BT-750 Operator Manual 44
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
4) Make sure that the probe is set on the body part correctly in every 2 or 3 hours. If the state
of your skin is different, change the part to be attached to other side.
Note
When probe is used incorrectly, inaccurate value or waveform can
come out as below.
● When the patient uses the probes without certification of Bistos
Technology.
● Hemoglobin dysfunction
Caution
Probe should be treated with attention. Careless use might cause
damage to probe. Keep the probe cable away from acute objects.
When the patient has a high fever or peripheral circulatory failure,
the skin, temperature increases to 2~3 degree.
In case of the patients with cold fingers, SpO2 data is not accurate.
For the patients who have abnormally high oxyhemoglobin or
methemoglobin , SpO2 data is not accurate.
You can not use disposable probe to different patients. For same
patient, it is possible to reuse and remeasure even in different
portion. If you want to reuse disposable probe, disinfection is
indispensable.
Caution
It might fail to measure pulse rate under occasions as follows;
When probe is fastened tightly.
When probe excessively exposes to Medical illumination, bilirubin
illumination or sunlight.
In case of being measured on arms or legs in cuffs or bandages.
Warning
In case of measuring with probe while MRI is in use, the patient may
have burns. To minimize the danger, use non-inductive line, if there
exists any danger even with proper use, remove the connector from
the patient.
● Allergic patients should not use disposable probes.
● Make sure that the probes radiate light, faced each other correctly,
and the light reaches to the tissue of a patient. When applying to a
newborn baby, use disposable probes outside of the incubator.
● If used in an incubator, accurate results may not be obtained
because the humid air can have an effect on the outcome.
When probe is excessively fixed with tape or used in other incorrect
way, the patient may suffer from injury.
Warning
If the probe are excessively exposed to medical illumination
( especially for xenon light ), bilirubin illumination, fluorescent
lamp, IR heating appliances or direct ray of light , this may cause
problem in function . To avoid excessive exposure to the light, use it
by the general operation and be careful that the probe is not exposed
to outside ray with opacity. If outdoor circumstance is under
extremely high light, it may generate inaccurate value.
Do not use probe at the body part where intravenous injection or
artery catheters are placed.
Do not use damaged SpO2 probe or optical appliances.
Do not clean with oxygenated water, ultraviolet ray, direct ray of
light and steam.