116
| Chapter 6: Error Codes
Overview
An error code that appears on the display of the ELx808 (during instrument control of the
reader), or in Gen5 software (during PC control of the reader) is displayed as a four-digit
identifier. The first digit will be 0, 1, 2, or A.
•
0, 1, or 2 denote a noncritical or “General” error, which means that it is still
possible for the ELx808 to communicate with the controlling software and run
the reader system test. See
beginning on page 117.
•
A denotes a more serious error with the memory or processing, which requires
that the ELx808 be turned off and then powered up before any diagnostics can
be performed. If the error reappears, call BioTek TAC for troubleshooting
assistance (see Chapter 1). See
Diagnostics
If an error code is displayed, a System Self-Test should be conducted for diagnostic
purposes. Refer to the following
brief
instructions for performing the Self-Test:
•
From the Main Menu screen on the ELx808 display, select
UTIL > TESTS
> SYSTEM
.
•
In Gen5, access the System menu and select
Diagnostics > Run System
Test
.
Terms Used in the Error Codes Tables
•
Air Blank:
A full light reading through a filter with no plate in the light path.
•
Dark Current:
A reading taken with the light blocked to measure background
light levels in the reading chamber. Also used as a measure of background
electronic noise within the measurement circuit.
•
Gain:
An automatic electronic adjustment to the measurement circuit. The
gain adjustment compensates for changing light levels or filter variations. For
example, if the lamp output decreases slightly, the gain will increase to make
up the difference.
•
Axis:
Refers to a motor that positions the filter wheel or plate carrier.
•
Offset:
A numerical limit, usually a range. For example, if the gain fails an
offset test, it may be too high or too low.
BioTek Instruments, Inc.
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...