36
| Chapter 3: Operation
MEAS Selection
The
MEAS
selection screen allows you to select the filter(s) for the assay. The
currently selected filter appears on the top line of the display, and the available
options appear on the bottom.
M E A S : 4 5 0
R E F : 6 3 0
4 0 5 4 5 0
4 9 0
6 3 0
•
Press
SOFT KEYS 1
,
2
,
3
,
and
4
to select the filter option displayed
above the soft key. The display updates to reflect the selection.
•
Press the right arrow key to select the
Reference Filter
.
•
Press
ENTER
to move to the next screen.
Number of Kinetic Reads/Kinetic Duration Selection
This menu allows you to either select the total number of kinetic reads or the
length of time the assay will run (kinetic duration). Any previously defined
value is shown on the top line of the display and the options on the second.
K I N E T I C :
T O T A L
R E A D S
T O T A L
R E A D S
D U R A T I O N
•
Press
SOFT KEY 1
to select the total reads option.
•
Press
SOFT KEY 3
to select the duration option.
•
Press
Enter
to save the selection and advance to the next screen.
Kinetic Interval
Use this screen to enter the interval of time between each kinetic read.
K
I
N
E
T
I
C
I N T E R V A L :
0 1 : 2 3 : 5 6
•
Use the
NUMERIC
keys to enter the time duration. Valid ranges are: 0-1
hours, 0-59 minutes and 0-59 seconds. The number of Reads =
Duration/Interval must be less than or equal to 40 and more than or
equal to 2.
BioTek Instruments, Inc.
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...