46
| Chapter 3: Operation
Constant Blank Value Entry
This entry screen only appears when a Constant Blank map is selected. Enter
the value to be subtracted from each well on the plate.
E
N
T
E
R
B L A N K I N G
C O N S T A N T :
0 . 0 0
0
•
Use the
NUMERIC
keys to enter the value. The range is 0.000 to 3.000.
The cursor is positioned at the first editable field and advances
automatically.
•
Press the
CLEAR
key to clear the value on the display.
•
Press the
ENTER
key to save the value and advance.
Number of Blanks
The Number of Blanks field allows you to enter the number of blank wells in
the assay. This entry screen is only displayed when Full, Column, or Row
blank map is selected.
E N T E R N U M B E R
O F
B
L
A
N
K
S
: 0
2
•
Use the
NUMERIC
keys to enter the number of blanks. The range is 0 to
48.
•
Press the
CLEAR
key to clear the Number of Blanks value from the
display.
•
Press the
ENTER
key to save the value and move to the next screen.
Selecting a Blank Location
The Blank Location screen allows you to define where the blank well is located
on the microplate. This screen only appears if manual map generation has been
selected. Any previously defined value is displayed.
E N T E
R
T H E
L O C A T I O N
O F
B
L
A
N
K
#
1
:
A
1
2
•
Use the
NUMERIC
and
ALPHA
keys to enter a Blank Location, based
upon the plate geometry.
BioTek Instruments, Inc.
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...