Decontamination
|
111
Decontamination
Purpose
Any laboratory instrument that has been used for research or clinical analysis is
considered a biohazard and requires
decontamination
prior to handling.
Decontamination minimizes the risk to all who come in contact with the instrument
during shipping, handling, and servicing. Decontamination is required by the
U.S.
Department of Transportation
regulations.
Persons performing the decontamination process must be familiar with the basic setup
and operation of the instrument.
Important!
BioTek Instruments, Inc. recommends the use of
the following decontamination solutions and methods based on
our knowledge of the instrument and recommendations of the
Centers for Disease Control and Prevention (CDC). Neither
BioTek nor the CDC assumes any liability for the adequacy of
these solutions and methods. Each laboratory must ensure that
decontamination procedures are adequate for the Biohazard(s)
they handle.
Warning! Internal Voltage.
Turn off and unplug the
instrument for the decontamination procedure.
Warning!
Wear prophylactic gloves when handling
contaminated instruments. Gloved hands should be
considered contaminated at all times; keep gloved hands away
from eyes, mouth, nose, and ears. Eating or drinking while
decontaminating instruments is not advised.
Warning!
Mucous membranes are considered prime entry
routes for infectious agents. Wear eye protection and a surgical
mask when there is a possibility of aerosol contamination.
Intact skin is generally considered an effective barrier against
infectious organisms; however, small abrasions and cuts may
not always be visible. Wear protective gloves when performing
the Decontamination procedure.
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...