Sample Anticardiolipin IgG Enzyme Immunoassay Kit |
139
Criteria Range
(GPLU/mL)
High Positive
> 80
Medium Positive
≥
20-80
Low Positive
≥
10, < 20
Results Interpretation
The association between low positive levels of anticardiolipin antibodies and clinical
findings is unclear.
Normal population studies indicate that there is a higher prevalence of IgM positives
in the normal population than IgG, 9.4% and 6.5%, respectively. In normal pregnancy,
the levels are higher still at 17.0% (IgM) and 10.6% (IgG).
Expected Values
The normal range was determined on serum from 102 normal adult blood donors. The
ranges below are provided as a guide only. ELISA assays are very sensitive and
capable of detecting small differences in sample populations. It is recommended that
each laboratory determine its own normal range, based on the population techniques
and equipment employed.
lgG Anticardiolipin
< 11 GPL U/mL
Negative result
> 11 GPL U/mL
Positive result
Programming the Assay on the ELx808
From the Main Menu, press
DEFINE
. Enter the assay number and edit the name if
desired. At the
DEFINE
menu, follow the steps below:
STEP
To program the reading method, press:
METHOD:
READ TYPE: Endpoint
DELAY FIRST READ:
If incubator model: INCUBATION TEMP (Ambient or Temperature):
DELAY FIRST READ:
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...