Defining Assays
|
61
Transformation Formulas
Transformation formulas change the absorbance data of all wells defined in the
Map to another format, in preparation for further data reduction.
Transformation Formula Definition
•
From the assay
Define
Menu, press the arrow corresponding to
Formula
.
D
E
F
I
N
E :
M E T H O D
M A P
F O R M U L A
C U R V E
Ï
•
This will bring you to a screen asking to
Select Formula Type
. At this
screen, select
TRANS
and then enter the formula using the
Math
,
Other
,
Map
and
Function
keys.
S E L E C T
F O R M U L A
T Y P E :
C U T O F F
T R A N S
V A L
* M O R E
Ï
Example:
Divide all ODs on the plate by 2 and multiply by 100.
Enter the formula:
(OD/2)*100
This formula will be applied to the ODs of all samples, standards,
controls, and blanks that are present on the plate map.
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...