| Chapter 3: Operation
Examples
1:
The cutoff between negative and positive calls should be calculated as the
average of the negative controls plus the OD value of 0.500. Samples greater than
the cutoff should be labeled as positive. No greyzone is required.
•
For this example, NC;x (the mean of the NC wells) equals 1.000 OD
•
The cutoff formula is
NC;x+0.5
•
The greyzone is
00%
•
POS is selected for
SAMP>00%
•
Calls are assigned to sample wells as follows:
¾
EQUIV if the sample equals 1.500
¾
POS if the sample is greater than 1.500
¾
NEG if the sample is less than 1.500
2:
For a quantitative assay, samples with OD values greater than the STD2 mean
plus a 10% greyzone should be labeled as positive; samples with OD values less
than the STD2 mean minus the 10% greyzone should be labeled as negative. All
other samples should be considered equivocal.
•
For this example, STD2;x (the mean of the STD2 wells) equals 2.000 OD
•
The cutoff formula is simply
STD2;x
•
The greyzone is
10%
•
POS is selected for
SAMP>10%
•
Calls are assigned to sample wells as follows:
¾
EQUIV if the sample is greater than or equal to 1.800
and
less than or equal to 2.200
¾
POS if the sample is greater than 2.200
¾
NEG if the sample is less than 1.800
BioTek Instruments, Inc.
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...