Defining Assays
|
69
•
For example, to include Assays 1, 8, and 22 in the Panel assay (
DOWN
mapping is selected for the Panel):
¾
Assay 1 has a total of 12 wells defined for controls, blanks, and
standards. In the Panel, the mapping for Assay 1 begins in well A01. The
user wants to run 6 samples in Assay 1. Assay 1 now fills wells A01
through B03.
¾
The mapping for Assay 8 can begin in well B04, or any well other than
A01 to B03. The reader will “beep” if you try to map into a well that is
already assigned for use with the Panel.
¾
The mapping for Assay 22 may begin at the next available well location
after Assay 8 mapping is complete.
¾
The Panel Assay results are sorted by Sample (unless a custom assay has
been programmed by BioTek).
Note:
The Interpretation of Results reports for each assay in
the Panel will print first, and then the Sample results will print.
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...