Newer-Style Module |
143
Newer-Style Module
Adjust the Power Input Voltage Setting
Important
!
Turn off and unplug the reader before adjusting
the voltage setting.
1.
Unplug the reader and remove the power cord.
2.
Insert a small pair of
needle-nose pliers
into the two holes on the cover of
the power input module. Lift the cover open and swing it back toward the
power cord socket (see
below).
3.
As shown below, a small voltage selector card is located to the right of the fuse
holder. Using the pliers, carefully pull this card straight out of the
compartment.
Figure 12:
Power input module: accessing the voltage selector card and fuse holder
ELx808 Operator’s Manual
Summary of Contents for ELx808
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Page 2: ......
Page 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Page 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Page 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Page 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Page 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Page 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Page 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Page 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Page 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Page 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Page 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...