background image

15

PUMP SQUEEZE TUBE LUBRICATION AND 
REPLACEMENT

!

  Remove  and  inspect  the  squeeze  tube.  If  it  is 

round  and  not  flattened,  and  shows  no  visible 
signs of wear, then it can simply be lubricated as 
per the instructions in this section. 

The OxiBOSS monitors the running time of the dosing 
pump  and  will  display  a  reminder  to  replace  the 
squeeze tube when this time exceeds 50 hours.

If the squeeze tube appears a little flattened and 
has wear marks, then it is best to replace it. A new 
squeeze tube can be ordered  at bionizer.com.au.

!

  Switch off the OxiBOSS and ensure that the pool 

pump is also off. 

At this time, the peristaltic pump squeeze tube either 
needs  to  be  lubricated  or  replaced  as  per  the 
instructions below. 

Removing the Squeeze Tube

Damage  caused  by  chemical  leaking  from  a 
broken  squeeze  tube  is  not  covered  by  the 
warranty, so it is your responsibility to check and 
replace it as necessary.

Remove  the  Squeeze 
tube.  The  rotor  may 
come  out,  but  can  be 
pushed back in place.

P u l l   o ff   t h e   p u m p 
cover.

H o l d   t h e   t o p   a n d 
bottom  of  the  pump 
cover.

Lift up the pump cover 
from the bottom side.

MAINTENANCE

Apply Silicon

grease here.

Pull  out  the  pump  rotor  a  little,  and  place  the 
squeeze tube between the pump body and the two 
rollers. 

Push the squeeze tube and rotor into the pump body. 
Ensure  that  the  inlet  and  outlet  are  seated  in  the 
guides on each side of the pump body.

Refitting the Pump Cover

Re-fit the pump cover until it clips into place. There 
should  be  no  gaps  anywhere  between  the  pump 
body and cover. Ensure that the squeeze tube is not 
pinched on either side, after the cover is replaced.

Refitting the Squeeze Tube

Lubricating the Squeeze Tube

DO NOT 

over-lubricate, as this will attract dirt and 

grit, which will cause premature wear of the squeeze 
tube and rollers.

Apply    a 

THIN  SMEAR 

of  Silicon  grease  to  the 

squeeze tube and pump rotor shaft as shown above.

(Skip this step when replacing squeeze tube, as these 
are supplied pre-lubricated.)

WARNING : Use only silicon grease. Petroleum gel 
and  other  oil-based  lubricants  may  cause  the 
squeeze tube to break down, resulting in corrosive 
chemical spillage.

Summary of Contents for OxiBOSS

Page 1: ...d before you call us for advice Version 2 1a 19 Sep 2019 bionizer com au bionizer com Bionizer highly recommends that a pH BOSS automatic pH Controller is used together with the OxiBOSS Controller Whe...

Page 2: ...NG 17 BIONIZER CONTACT INFORMATION 20 CONTROL BOX INSTALLATION 9 MAINTENANCE 13 PUMP SQUEEZE TUBE LUBRICATION AND REPLACEMENT 15 TEE PIECE INSTALLATION 7 TEE PIECE HOSE INSTALLATION 10 AUTO OFF BUTTON...

Page 3: ...her load would need around 4ppM of Chlorine to maintain a chemical activity of 650 mV ORP A pool that is not being used during winter with low water temperature may need 1ppM of Chlorine to maintain t...

Page 4: ...omaticsanitisercontrol Presstoscrollthroughthesixusersettingmenus Presstosavechangeswhenadjustingsettings 16x2Linealphanumeric LCD with LED backlight Presstomakeschangestousersettings Precision high q...

Page 5: ...Main OxiBOSS Display 5 ThecurrentORPofthePool The Set Point at which the OxiBOSS will maintaintheORPlevel Seepage12 The arrow is also animated whenever thedosingpumpisaddingchemical The Arrow here ind...

Page 6: ...nd using this electrical equipment includingthefollowing 5 Always power the OxiBOSS from a genuine Bionizer power adaptor An incorrect adaptor could damage the OxiBOSS and will void your warranty 5 Re...

Page 7: ...section of pipe sufficient for the Probe Injection Tee piece For 40mm 1 5 pipe where the reducing busheswillbeused cutouta7cm 2 75 section ofpipe NOTE Turn off all valves prior to fitting the T piece...

Page 8: ...rom vertical DO NOT FIT THE TEE TO A VERTICAL PIPE AS THIS WILL MAKE THE ORP PROBE HORIZONTAL WHICH WILL RENDER THE ORP PROBE INOPERABLE Correctly Installed Tee Incorrectly Installed Tee Do not mount...

Page 9: ...resistant Any water ingress due to the unit being incorrectlymountedisnotcoveredbywarranty 1 Choose a position that will allow you to see the OxiBOSS Display easily Fix the control box in positionusi...

Page 10: ...y by hand A tool may be used for another to 1 turn if the fitting is leaking when the pumpisrunning DO NOT OVER TIGHTEN 3 After the ORP Probe is installed in the Tee piece connect plug on the ORP Prob...

Page 11: ...water This is related tobothpoolsize andpump pipesetup In general a smaller pool would use a lower Dose Setting than a large pool However other factors can affectthis suchaspoolturnovertime 20 250 000...

Page 12: ...THE SANITISER PUMP WILL BE DISABLED AND THE OxiBOSS WILL NOT PUMP MORE SANITISER UNTIL THE ALARM HAS BEEN RESET Problem DoseSettingistoolowortoohigh Fix See page 11 for information on how to changethe...

Page 13: ...rnalpowersource pleaseensurethat the pool pump is switched off and unplugged duringtheentirecalibrationprocedure The ORP Probe should be calibrated to ensure that the readout is correct Calibration is...

Page 14: ...t into the Tee Re connect the ORP Probe to OxiBOSS when done Switchyourpoolequipmentbackon Re connect the pool pump to the Bionizer power outlet or the power output to which it was connectedbeforestar...

Page 15: ...an be pushedbackinplace Pull off the pump cover Hold the top and bottom of the pump cover Lift up the pump cover fromthebottomside MAINTENANCE Apply Silicon grease here Pull out the pump rotor a littl...

Page 16: ...her lubricated or replaced the last step is to reset the Squeeze Tube remindertozero Replacement Squeeze Tubes are supplied with new Supply and Delivery tubes pre fitted for your convenience It will b...

Page 17: ...ntactyourdistributororBionizer Faulty AC DC poweradaptor FaultyOxiBOSS mainunit ReturntoBionizerforrepair ReducetheDoseSettingbyonelevelandcheckin2days DoseSettingissettoohigh Replacesqueezetubeand or...

Page 18: ...ayed when the peristaltic dosing pump has exceeded 50 hours of operation since the last time the Squeeze Tube was lubricated or changed ThismessageisdisplayedeachtimetheOxiBOSSisswitchedon If you need...

Page 19: ...s Reduces Copper Ion loss from various environmental stressors Use a double dose of Ionlife when going away for extended holidays to help ensure that you return to a pristinepool MayextendthelifeofCop...

Page 20: ...Email relax bionizer com 16427 N Scottsdale Rd Suite 410 USA Toll Free 1 855 8 IONIZE Web bionizer com QLD 4558 Australia Web bionizer com au 1 173 Bradman Ave Phone 1800 10 62 74 Fax 61 7 5443 6700 E...

Reviews: