 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

MAINTENANCE

A LITTLE PREVENTATIVE MAINTENANCE WILL HELP YOU GET THE MOST FROM YOUR OxiBOSS

MAINTENANCE

13

30

20

10

100

50

!

There are TWO bottles of 
ORP  Calibration  solution 
(ZoBell Part A & Part B).

!

 Empty the entire contents 
of small Part B bottle into 
the larger Part A bottle.

!

 Replace  the  lid  onto  the 
Part  A  bottle  a  shake 
gently to mix the contents 
well.

!

Dispose  of  solution  in 
normal  drain  such  as  a 
kitchen  sink.  Solution 
expires within 4 weeks of 
being made up. 

!

 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

100

50

!

 Leave  the  ORP  Probe  in  this 
ORP  Standard  for  at  least  10 
seconds  before  proceeding 
with  the  calibration  steps 
below.

!

Place the ORP Probe into ORP 
Standard,  as  far  as  the  probe 
will go into the neck.

!

 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

!

 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

!

 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

!

 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

!

 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

!

 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

!

 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

!

 Unplug  the  ORP  Probe  from  the  OxiBOSS  first. 
(This will avoid the cable getting twisted when the 
ORP Probe is unscrewed.  U n s c re w   t h e   p ro b e 
completely from the Tee.

!

 When  the  reading  is  stable  and  the  “

ü

”  is 

displayed, press 

PROGRAM / SET

 to calibrate.

5. Calibration Procedure

!

 Rinse the probe well in clean water. De-ionized or 
distilled water is ideal, but tap water will suffice.

4. Place OxiProbe into 230mV ORP Standard

!

 Press the 

PROGRAM / SET 

button until the ORP 

Calibration screen is displayed, as per the example 
below.  The  “

Current

”  reading  may  be  220  or 

another value, depending on the condition of the 
ORP Probe and current calibration setting.

!

 Reconnect  the  ORP  Probe  connector  to  the 
OxiBOSS.

!

 Adjust  the  “

Cal  ORP

”  value  to  exactly  230.  The 

flashing “

X

”, indicates that the reading is not yet 

stable. DO NOT calibrate until the “

X

” changes to a 

ü

”.

3. Remove the ORP Probe from the Tee

!

 Switch the OxiBOSS power on. If it is connected to 
a Bionizer, set the Bionizer to Manual:On mode to 
switch on the power outlet.

!

  Press the 

PROGRAM / SET 

button twice to bypass 

the start-up screen delay.

!

 If you are using a Bionizer to control the power to 
your  pool  pump  and  OxiBOSS,  unplug  the  pool 
pump from the Bionizer Control Unit.

ORP PROBE CALIBRATION

he OxiBOSS will display a reminder to calibrate the 
ORP Probe approximately every 6 months:

!

 (Connect the OxiBOSS to a power outlet socket, 
but do not switch on yet.)

!

 (If you are powering the OxiBOSS and Pool Pump 
from an external power source, please ensure that 
the  pool  pump  is  switched  off  and  unplugged 
during the entire calibration procedure.)

The ORP Probe should be calibrated to ensure that 
the  readout  is  correct.  Calibration  is  simply  the 
process of placing the ORP Probe into a solution of 
known ORP, and adjusting the readout to that value.

2. Prepare 230mV ORP Calibration Standard

!

 Leave the OxiBOSS connected to the Bionizer.

1. Prepare for OxiBOSS for Calibration

100

50

Summary of Contents for OxiBOSS

Page 1: ...d before you call us for advice Version 2 1a 19 Sep 2019 bionizer com au bionizer com Bionizer highly recommends that a pH BOSS automatic pH Controller is used together with the OxiBOSS Controller Whe...

Page 2: ...NG 17 BIONIZER CONTACT INFORMATION 20 CONTROL BOX INSTALLATION 9 MAINTENANCE 13 PUMP SQUEEZE TUBE LUBRICATION AND REPLACEMENT 15 TEE PIECE INSTALLATION 7 TEE PIECE HOSE INSTALLATION 10 AUTO OFF BUTTON...

Page 3: ...her load would need around 4ppM of Chlorine to maintain a chemical activity of 650 mV ORP A pool that is not being used during winter with low water temperature may need 1ppM of Chlorine to maintain t...

Page 4: ...omaticsanitisercontrol Presstoscrollthroughthesixusersettingmenus Presstosavechangeswhenadjustingsettings 16x2Linealphanumeric LCD with LED backlight Presstomakeschangestousersettings Precision high q...

Page 5: ...Main OxiBOSS Display 5 ThecurrentORPofthePool The Set Point at which the OxiBOSS will maintaintheORPlevel Seepage12 The arrow is also animated whenever thedosingpumpisaddingchemical The Arrow here ind...

Page 6: ...nd using this electrical equipment includingthefollowing 5 Always power the OxiBOSS from a genuine Bionizer power adaptor An incorrect adaptor could damage the OxiBOSS and will void your warranty 5 Re...

Page 7: ...section of pipe sufficient for the Probe Injection Tee piece For 40mm 1 5 pipe where the reducing busheswillbeused cutouta7cm 2 75 section ofpipe NOTE Turn off all valves prior to fitting the T piece...

Page 8: ...rom vertical DO NOT FIT THE TEE TO A VERTICAL PIPE AS THIS WILL MAKE THE ORP PROBE HORIZONTAL WHICH WILL RENDER THE ORP PROBE INOPERABLE Correctly Installed Tee Incorrectly Installed Tee Do not mount...

Page 9: ...resistant Any water ingress due to the unit being incorrectlymountedisnotcoveredbywarranty 1 Choose a position that will allow you to see the OxiBOSS Display easily Fix the control box in positionusi...

Page 10: ...y by hand A tool may be used for another to 1 turn if the fitting is leaking when the pumpisrunning DO NOT OVER TIGHTEN 3 After the ORP Probe is installed in the Tee piece connect plug on the ORP Prob...

Page 11: ...water This is related tobothpoolsize andpump pipesetup In general a smaller pool would use a lower Dose Setting than a large pool However other factors can affectthis suchaspoolturnovertime 20 250 000...

Page 12: ...THE SANITISER PUMP WILL BE DISABLED AND THE OxiBOSS WILL NOT PUMP MORE SANITISER UNTIL THE ALARM HAS BEEN RESET Problem DoseSettingistoolowortoohigh Fix See page 11 for information on how to changethe...

Page 13: ...rnalpowersource pleaseensurethat the pool pump is switched off and unplugged duringtheentirecalibrationprocedure The ORP Probe should be calibrated to ensure that the readout is correct Calibration is...

Page 14: ...t into the Tee Re connect the ORP Probe to OxiBOSS when done Switchyourpoolequipmentbackon Re connect the pool pump to the Bionizer power outlet or the power output to which it was connectedbeforestar...

Page 15: ...an be pushedbackinplace Pull off the pump cover Hold the top and bottom of the pump cover Lift up the pump cover fromthebottomside MAINTENANCE Apply Silicon grease here Pull out the pump rotor a littl...

Page 16: ...her lubricated or replaced the last step is to reset the Squeeze Tube remindertozero Replacement Squeeze Tubes are supplied with new Supply and Delivery tubes pre fitted for your convenience It will b...

Page 17: ...ntactyourdistributororBionizer Faulty AC DC poweradaptor FaultyOxiBOSS mainunit ReturntoBionizerforrepair ReducetheDoseSettingbyonelevelandcheckin2days DoseSettingissettoohigh Replacesqueezetubeand or...

Page 18: ...ayed when the peristaltic dosing pump has exceeded 50 hours of operation since the last time the Squeeze Tube was lubricated or changed ThismessageisdisplayedeachtimetheOxiBOSSisswitchedon If you need...

Page 19: ...s Reduces Copper Ion loss from various environmental stressors Use a double dose of Ionlife when going away for extended holidays to help ensure that you return to a pristinepool MayextendthelifeofCop...

Page 20: ...Email relax bionizer com 16427 N Scottsdale Rd Suite 410 USA Toll Free 1 855 8 IONIZE Web bionizer com QLD 4558 Australia Web bionizer com au 1 173 Bradman Ave Phone 1800 10 62 74 Fax 61 7 5443 6700 E...

Reviews: