Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
MAINTENANCE
A LITTLE PREVENTATIVE MAINTENANCE WILL HELP YOU GET THE MOST FROM YOUR OxiBOSS
MAINTENANCE
13
30
20
10
100
50
!
There are TWO bottles of
ORP Calibration solution
(ZoBell Part A & Part B).
!
Empty the entire contents
of small Part B bottle into
the larger Part A bottle.
!
Replace the lid onto the
Part A bottle a shake
gently to mix the contents
well.
!
Dispose of solution in
normal drain such as a
kitchen sink. Solution
expires within 4 weeks of
being made up.
!
Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
100
50
!
Leave the ORP Probe in this
ORP Standard for at least 10
seconds before proceeding
with the calibration steps
below.
!
Place the ORP Probe into ORP
Standard, as far as the probe
will go into the neck.
!
Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
!
Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
!
Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
!
Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
!
Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
!
Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
!
Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
!
Unplug the ORP Probe from the OxiBOSS first.
(This will avoid the cable getting twisted when the
ORP Probe is unscrewed. U n s c re w t h e p ro b e
completely from the Tee.
!
When the reading is stable and the “
ü
” is
displayed, press
PROGRAM / SET
to calibrate.
5. Calibration Procedure
!
Rinse the probe well in clean water. De-ionized or
distilled water is ideal, but tap water will suffice.
4. Place OxiProbe into 230mV ORP Standard
!
Press the
PROGRAM / SET
button until the ORP
Calibration screen is displayed, as per the example
below. The “
Current
” reading may be 220 or
another value, depending on the condition of the
ORP Probe and current calibration setting.
!
Reconnect the ORP Probe connector to the
OxiBOSS.
!
Adjust the “
Cal ORP
” value to exactly 230. The
flashing “
X
”, indicates that the reading is not yet
stable. DO NOT calibrate until the “
X
” changes to a
“
ü
”.
3. Remove the ORP Probe from the Tee
!
Switch the OxiBOSS power on. If it is connected to
a Bionizer, set the Bionizer to Manual:On mode to
switch on the power outlet.
!
Press the
PROGRAM / SET
button twice to bypass
the start-up screen delay.
!
If you are using a Bionizer to control the power to
your pool pump and OxiBOSS, unplug the pool
pump from the Bionizer Control Unit.
ORP PROBE CALIBRATION
he OxiBOSS will display a reminder to calibrate the
ORP Probe approximately every 6 months:
!
(Connect the OxiBOSS to a power outlet socket,
but do not switch on yet.)
!
(If you are powering the OxiBOSS and Pool Pump
from an external power source, please ensure that
the pool pump is switched off and unplugged
during the entire calibration procedure.)
The ORP Probe should be calibrated to ensure that
the readout is correct. Calibration is simply the
process of placing the ORP Probe into a solution of
known ORP, and adjusting the readout to that value.
2. Prepare 230mV ORP Calibration Standard
!
Leave the OxiBOSS connected to the Bionizer.
1. Prepare for OxiBOSS for Calibration
100