Maintenance and cleaning
Volume test
5-10
PREVI™ Color Gram
User’s Manual
V o l u m e t e s t
The maintenance kit is required for this test that must be performed
weekly
.
Press
Press
DECOLORIZER 3
FULL CAROUSEL
1 = PATTERN TEST
2 = VOLUME TEST
3 = B-LINE FLUSH
4 = PRIME LINES
VOLUME TEST
PRESS
ANY PRIME BUTTON
Hold a volume test
tube (small tube) in
place so that the
nozzle C is covered to
capture dispensed
liquid.
Press
briefly.
Remove and close the
tube.
Record the reagent
letter on the tube.