Maintenance and cleaning
Long-term storage
PREVI™ Color Gram
User’s Manual
5-17
Install a cap on the
reagent bottle and set
the bottle aside.
Flush at least 250 ml of
ethanol or methanol
through the line and
spray the nozzle by
priming.
Leave ethanol or
methanol in the line.
Remove and clean the
nozzles, see
"Disassembly",
page 5-4 and
"Cleaning", page 5-5.
IMPORTANT! Leave flushing fluids in the reagent lines during storage.
Do not run reagent lines dry.
Flush the bowl with
distilled water.
To operate the
instrument again,
perform the previous
steps in reverse order.
CAUTION! Do not subject the instrument to freezing temperatures if
aqueous fluids remain in any of the reagent lines.
Serious damage may result.