Maintenance and cleaning
Fuse replacement
PREVI™ Color Gram
User’s Manual
5-23
F u s e r e p l a c e m e n t
DANGER!
To prevent the risk of fire, the main fuses should only be replaced with fuses
of the same type and rating.
If fuse failure recurs, please call your bioMérieux representative.
CAUTION!
Fuse failure may indicate a serious internal problem. For continued protection,
replace only with fuses of the type and rating indicated on the rear panel
of the stainer.
Put the power switch
in position
O
.
Disconnect the power
cord from the mains.
Disconnect the main
power cord from the
socket.
Open the cover with a
small screwdriver.
Push the fuse holder
to the left to check the
fuses.
Replace fuses if
necessary.
Push the fuse holder
to the right.
Close the cover.
Reconnect the power
cord (socket and
mains) and switch on
the instrument.