Operating the PREVI™ Color Gram
CLEAN alcohol purge stage
4-10
PREVI™ Color Gram
User’s Manual
C L E A N a l c o h o l p u r g e s t a g e
CAUTION!
The CLEAN cycle must be used each time the PREVI™ Color Gram is left on
stand-by FOR MORE THAN 3 HOURS.
The alcohol purge stage must be performed after each staining period and
the prime stage must be performed just before restarting the instrument.
This stage requires the use of 10 ml of alcohol divided equally between each
nozzle.
Be sure the stainer
light is ON.
Place the empty
carousel in the
instrument and
close the lid.
Press .
Open the lid and
remove the carousel.
Switch OFF the
instrument.
Spray and wipe the
interior of the bowl
and the nozzles with
ethanol 70%.
IMPORTANT!
After this first cleaning stage, if the instrument remains on stand-by for more
than one week, please refer to the "Long-term storage", page 5-16.
DECOLORIZER 3
FULL CAROUSEL
PRESS
CLEAN TO REPRIME