Maintenance and cleaning
Volume test
PREVI™ Color Gram
User’s Manual
5-11
Place the tube in
position C.
Repeat steps 3, 4, 5
and 6 for all the other
nozzles
(D
F
, D
R
, A, B and E).
Check each tube
according to the
following table.
If the volume is
in the tolerance range,
go to point 9.
If the volume
obtained is out of the
tolerance range:
•
too low, loosen the
nozzle slightly
•
too high, tighten the
nozzle slightly
And repeat the test.
Tolerance
Tube
Minimum
Maximum
A
10 ml
12 ml
B, C
9 ml
11 ml
D
F
, D
R
, E
9 ml
If the volume is in the tolerance range, go to point 9.
If the volume is still out of the tolerance range,
see "Nozzle handling", page 5-4 to clean the nozzle.
Repeat the tests with the cleaned nozzles.
If the volume is in the tolerance range, go to point 9.
If the volume is still out of the tolerance range,
contact your bioMérieux representative.