Operating the PREVI™ Color Gram
Other settings
PREVI™ Color Gram
User’s Manual
4-13
E mp t yin g th e w ast e co nt ain er
WEEKLY PROCESS
Disconnect the waste
tubing from the full
waste container.
Empty the waste
container.
Throw the effluents
into an appropriate
biohazard bin in
accordance with local
regulations.
Connect the waste
tubing to the empty
waste container.
The instrument is ready
for use.
Note:
bioMérieux recommends emptying the waste container once a
week: see "Maintenance and cleaning", page 5-1.
O t h e r s e t t i n g s
WARNING!
The information contained in this section is intended for customers who
wish to use this equipment with applications other than those specified by
bioMérieux.
In that case, the user assumes complete responsibility for that use.
Deco lo ri zer settin g
The decolorizer cycle intensity can be changed according to the type of
specimens or the staining results expected.
IMPORTANT!
Levels 2 and 3 have been validated by bioMérieux (level 3: default setting).
Keys
DECOLORIZER CYCLE
ULTRA THIN SPECIMEN
ULTRA THICK SPECIMEN
LEVELS RECOMMENDED
BY BIOMERIEUX