Gene Q Thermal Cycler
Appendix 2 Parameter Value & Meaning
—
25
—
APPENDIX 2 Parameter Value & Meaning
Item
Parameter
Meaning
Range
Note
1
Temp(C)
Temp. Control point
4.0~99.0
2
Time
Time of Temp. Control
(
mm:ss
)
--:--
(
set the value before
“
:
”through “Function” key.
)
The time of Temp. control is
long limitlessly, meanwhile, the
hotlid closed.
00
:
00~59
:
59
3
Ramp(C/S)
Ramping Rate
(℃
/Second
)
0.1 ~5.0
#.#
Maximum of Ramping Rate
4
+Temp
Temp. Gain of Every
Cycler
0.0~9.9
5
+Time
Time Gain of Every
Cycler
(
mm:ss
)
0:00~9:59
6
Cycle1
Section Numbers
1~5
Cycle1
:
×00 From 00 to 00
×00
Cycler Numbers from
beginning to the end
1~99
From 00
The start segment of
the section
X
(
and existed segment
no.
)
to 00
The end segment of
the section.
≤ X+15
7
Grad.
Gradient Temp.
Difference
(℃)
0 ~30
87
Date
Date(YY-MM-DD)
00
(
~99
)
-01
(
~12
)
-01
(
~31
)
year-month-date
89
Time
Time(HH:MM:SS)
00:00:00~23:59:59
hour
:
minute
:
second
910
Default File
The default file name
Max. length is 11
characters.
1
10
Default User
The default user
name
Max. length is 11
characters.
1
12
Control Mode
Control Mode
Block
Under
this
mode
(once
reaching target temperature
namely
entering
into
the
thermostatical state), the time
for reagent’s temperature under
target temperature is less than
setting time (see Fig. 1).
Tube
Under this mode (once entering
into the thermostatical state
after
reaching
target
temperature and decreasing
about
3
℃
),
reagent’s
temperature can be close to
target temperature very fast
(see Fig. 2).
1
23
Sample Volume
Sample Volume
(
ul
)
10~200
1
34
Hotlid Temp
State of Hotlid: OFF
OFF
State of Hotlid: Open
(℃)
30~99
1
45
Block Temp
State of Block Temp
during the heating of
Hotlid
(℃)
0~99
1
56
Key Beep
Key Sound (ON or
OFF)
OFF/ON
Humming
1
67
Run End Beep
Run End Sound (ON
or OFF)
OFF/ON
Humming continuously
Summary of Contents for GeneQ Thermal Cycler
Page 33: ...NOTES...