Gene Q Thermal Cycler
Chapter 4 Operation Guide
—
14
—
When the file is normally running, “
” at
“Now Running” will continuously flash.
When running is over, the system will
remind “File running is over ,Press ‘Enter’
back to the main interface.”.
Press “
Stop
” and it will remind “Do you
want to stop running? Please select
‘F1’ or ‘F3’”.According to the reminding
information to select “Stop” or “Start”.
Press “
Pause
” and it will remind
“Pause state now. Press ‘F3’ to
continue running.” According to the
reminding information you can select
“Strat” to continue.
Press “
View File
” to view edited PCR
programs.
Press
“
Skip
”
to
stop
current
temperature and run the next.
3
How to Set System Parameters
In the main interface press “
System
” to
enter system parameter setting interface.
Press “
Enter
” to move the cursor to next
item. The option is followed by the symbol
“ ”. Press “
Up/Down
” to roll the option for
selecting, and other information can be
entered by pressing “
Digit & Letter
”. Press
“
Right/Left”
to move cursor’s position.
Press “
Back
” to go back to the main
menu.
Press “
Save
” to save parameters.
Special
Reminding:
When
set
the
parameter,
please
press
the
“Save”,
otherwise
the
parameter
remains
the
original one.
3.1
“Date”and“Time”stand for default time after power-on.
3.2
“Default File” and “Default User” stand for name and user of default running procedure after
power-on.
3.3
“Module Select” stands for module type, including two optional modes, named “18*0.5” and
“24*0.2”. Selecting the module type according to the real situation.
3.4
“Control Mode” stands for default control model after power-on, including two optional modes,
named “Block” and “Tube”.
“Block” mode
subjects to simple temperature control during the course of cooling & heating
only. Compared to the block, cooling & heating rate of reagent is a little slower, therefore, the
actual change of reagent’s temperature is different from the setting procedure. The actual time
for reagent to reach & keep the setting temperature is far shorter than the setting time.
File running interface
System Parameter Setting Interface
Hotlid=off
2007-11-1 10:05:00
100
75
50
25
0
Temp
1c 95.0
2c 55.0
3c 72.0
4 04.0
Time
1:00
02:00
01:45
00:00
File Name:PCR1
User: BIOER
Now Running
1c
Temp=00.0
Time=00:00
Cycle 01 of 10
Total Time: 00h 56m
Remainig Time:00h 56m
Stop
Pause
Start
Skip
View File
Hotlid=off
2007-07-28 10:06:00
Date(yy-mm-dd):
07-07-28
Time(hh-mm-ss):
10:06:19
Default User:
BIOER
Default File:
PCR1
Module Select:
24*0.2
Control Mode:
Block
Hotlid Temp:
OFF
Block Temp:
10C
Key Sound:
OFF
Run End Sound:
ON
Back
Save