Gene Q Thermal Cycler

Chapter 4 Operation Guide

11

Press “

Edit

” to edit the file indicated by

the cursor.

Press “

New File

”to edit a new file.

Press “

Delete

” and the system will alarm

“Do you want to delete the file? Please
select

F3

or ‘F4’”, then according to

the alert to select

Delete

or

Back

.

Press “

Back

” to go back to the main

interface.

If the file has no password, press “

Edit

” to

enter file editing interface; While if the file
has password, press “

Edit

”, thus a dialog

will be displayed to indicate the user to
enter the password.

Press

Character button

in the main

interface to input the password directly.
Press “

Back

” to go back file list interface.

Press

“Accept

” to confirm the password,

and then the system will enter the file
editing interface,

If the file has no password, press “

Run

” to

enter file running interface or sample
volume input interface. If the file has
password, press “

Run

”, a dialog box will be displayed for inputting the password. Press

“Back

to go back the file list interface. Press “

Accept

” to confirm the password, and then the system

will enter file running interface of sample volume input window.

After press “Run” in the file list interface, and
input the password(If the file has Password.):

If the current Control Mode is Block mode,
Press “

Accept

” to enter file running

interface; while if the current Control Mode
is Tube mode, press “

Accept

”, thus a

dialog box will be displayed for inputting
sample volume.

Press

Character button

to input sample

volume according to practical condition.
Press “Back” to go back file list interface.
Press “Run” to enter file running interface.

Hotlid=off

2008-07-28 10:05:00

No.

User

File Name

Save Time

1

BIOER1*

PCR1

2007-07-28

2

BIOER2

PCR2*

2007-07-28

3

BIOER3

PCR3

2007-07-28

Back

Accept

Enter password
-

Please enter sample volume:

020μl

Enter password
-

Hotlid=off

2008-07-28 10:05:00

No.

User

File Name

Save Time

1

BIOER1*

PCR1

2008-07-28

2

BIOER2

PCR2*

2008-07-28

3

BIOER3

PCR3

2008-07-28

Edit

New File Delete

Back

Run

Hotlid=off

2008-07-28 10:05:00

No.

User

File Name

Save Time

1

BIOER1*

PCR1

2008-07-28

2

BIOER2

PCR2*

2008-07-28

3

BIOER3

PCR3

2008-07-28

Back

Run

Please enter sample volume:

020μl

Flie List Interface

Summary of Contents for GeneQ Thermal Cycler

Page 1: ...art of this manual shall be photocopied reproduced or translated into other languages without written permission from the Bioer Co Patent No ZL 98 2 44367 6 Standard No YZB 0189 2008 Document Number B...

Page 2: ...cler s power line is a safety device that should be matched with a grounded socket Never let the third ground pin floating If the 3 pin plug cannot be inserted it is recommended to ask an electrician...

Page 3: ...hould be protected against any kind of heat sources like sunlight ovens or central heating equipment If the Thermal Cycler is set aside for a long time it is recommended to disconnect the power line t...

Page 4: ...hole and the tube inside it In case it is smeared the surface of the Thermal Cycler may be scrubbed with a piece of dehydrated soft wet cloth 7 WARRANTY AND SERVICE INFORMATION a Warranty The Thermal...

Page 5: ...cation and unauthorized maintenance or alteration The Bioer Co makes no express warranties other than those which are described herein Any descriptions in sales promotion under specific conditions sha...

Page 6: ...ption 6 2 Keyboard Diagram 7 3 Key Functions 7 4 Inspection before Power on 8 5 Power on Procedures 8 CHAPTER 4 OPERATION GUIDE 9 1 How to Run a PCR Program 9 2 How to edit a PCR program 10 3 How to S...

Page 7: ...n and adjustment of DNA sequence of protein structure detection of gene extension construction of synthetic genes construction of cloning and expression carrier detection of polymorphism of a gene s i...

Page 8: ...ree water baths cyclically and thus finishes the three processes of denaturation annealing and extension This device has high temperature control accuracy high heating cooling rate and excellent tempe...

Page 9: ...g thermal cyclers have gradually dominated the market 5 Features of the Gene Q Thermal Cycler The Gene Q Thermal Cycler is a thermoelectric cooling thermal cycler It features a Reliable and stable ope...

Page 10: ...20 240V 50 60Hz 200W 100 120V 50 60Hz 200W Note Before power on please check if the above operating conditions are satisfied Pay special attention to the power line s reliable grounding 2 Transportati...

Page 11: ...situ time 99 hour 59 min Programs storage 100 5 Software Functions File editing and saving File accessing modifying and deleting Automatic file link Heating cooling ramp setting Automatic temperature...

Page 12: ...Description Heated lid Block Block LCD display Keyboard RS232 interface Power switch Power connecter Fuse Socket Note 1 The fuse specifications are as follows 250V 2 5A or 125V 5A 5 20mm The fuse shou...

Page 13: ...e left or right when pressing during file parameter setting Up Down moving the cursor one position to the up or down when pressing during file parameter setting Enter press it to accept the present pa...

Page 14: ...nding of the power line is reliable 5 Power on Procedures Turn on the power i e turn the power switch to position 1 the Thermal Cycler will beep twice and the LCD will display self test That means the...

Page 15: ...evolve the knob clockwise till it sound dang dang Revolve the knob one more turn At the beginning of revolving the knob clockwise or counter clockwise you can hear dang dang sound press the release ke...

Page 16: ...o input sample volume according to practical condition Under Tube mode the overshooting temperature and keeping time will be dirrerent in the case of different sample volume Press Back to go back the...

Page 17: ...terface Press Accept to confirm the password and then the system will enter file running interface of sample volume input window After press Run in the file list interface and input the password If th...

Page 18: ...ess Seg to enter segment edit interface In every segment you can set setting temperature Temp duration Time Ramping rate Ramp temperature increment of every cycler Temp and time increment of every cyc...

Page 19: ...ccording to practical condition Press Back to go back file save interface Press Run to enter file running interface Enther the file save interface if the user press Run without pressing Save the syste...

Page 20: ...nal one 3 1 Date and Time stand for default time after power on 3 2 Default File and Default User stand for name and user of default running procedure after power on 3 3 Module Select stands for modul...

Page 21: ...emperature which can allow reagent to reach the setting temperature faster During the same setting time as Block the actual time for reagent to keep the setting temperature is prolonged obviously Plea...

Page 22: ...Upgrading Software 1 Choose and click Setup exe in the CD start up guide procedure and then press Next to continue installation refer to the picture below 2 Please confirm License Agreement of the so...

Page 23: ...xt to enter into the installation interface 4 Please select target installation folder refer to the following picture and then press Next to enter into the installation interface The guide procedure w...

Page 24: ...fy and press Next to edit installation subassembly press Finish to finish this step refer to the following picture Note It is necessary to select Installation file If you delete before installation fi...

Page 25: ...ade exe under the folder of upgrading software installation the screen will show you the upgrading procedure interface refer to Figure 1 Press Expand to enter into the expansion interface refer to Fig...

Page 26: ...Open the control panel of computer to click Device Manager and then double click COM1 as showed in Figure 3 2 Choose Advanced in Communications Port COM1 Properties Revise the COM Port Number in Adva...

Page 27: ...elow ambient one Cooling module failure Contact distributor or manufacturer Fan failed or stopped 4 Heating and cooling terminated Bad sensor Contact distributor or manufacturer Malfunction of all coo...

Page 28: ...ributor with Heatsink temperature is too high Heat sink temp higher than 70 the sensor broken or short circuit or the fan is not working 3 Error Message when running Please contact the distributor wit...

Page 29: ...1_SEN LID J2 2_SEN LID J2 3_HEATER LID J2 4_HEATER CPU PCB4 BY Q5041041000000_CPU J13 1_HV J13 2_HV J13 3_VIN5 J13 4_HOTLID J13 5_FAN J13 6_PWM J13 7_HEAT J13 8_COOL J3_CPU_JTAG_IDC20 J10_KEY IN J5_LC...

Page 30: ...TAG_IDC20 J10_KEY IN J5_LCD_24P J9_CPLD_JTAG_IDC10 J16 1_SEN HOTLID J16 2_SEN HOTLID J15 1_SEN FAN J15 2_SEN FAN J15 3_SEN BLOCK J15 4_SEN BLOCK VB1_SUB D9_RS232C J17 1_H J17 2_H L U5 1 2 3 4 5 6 LIN...

Page 31: ...00 99 01 12 01 31 year month date 89 Time Time HH MM SS 00 00 00 23 59 59 hour minute second 910 Default File The default file name Max length is 11 characters 110 Default User The default user name...

Page 32: ...Gene Q Thermal Cycler Appendix 2 Parameter Value Meaning 26 Picture 2 Picture1 block reagent block reagent...

Page 33: ...NOTES...

Reviews: