9.3 Installation of the voltage and frequency changer
(chambers with voltage and frequency changer)
The voltage and frequency changer is supplied packed separately together with the battery test chamber.
CAUTION
Risk of injury and danger of damages by lifting heavy loads and by sliding or tilting
of the voltage and frequency changer in case of improper lifting.
Injuries; damage to the voltage and frequency changer.
∅
Do NOT lift the voltage and frequency changer by hand.
Lift the voltage and frequency changer from the pallet using technical devices (fork
lifter). Set the fork lifter only from the rear in the middle of the chamber.
Alternatively, the voltage and frequency changer can also be lifted at the eyelets on
the top by means of a lifting crane or fork lifter
(a) Eyelets for lifting with a lifting crane or fork lifter
(b) Positions for a fork lifter
Figure 18: Positioning of aids for lifting the voltage
and frequency changer
For the installation of the voltage and frequency changer behind the battery test chamber, provide a rear
wall distance of the chamber of approx. 1 m /
3.3 ft
.
If possible, fix the voltage and frequency changer at the battery test chamber. For this purpose, an Allen
key size 4 is required. Connect the slots at the end of the chassis with two M6 screws to the threads
provided below on the rear panel of the battery test chamber (see Figure 19).
(a)
(b)
(a)
(b)
Summary of Contents for 9020-0402
Page 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Page 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Page 162: ...MK LIT E5 01 2021 page 162 172...
Page 163: ...MK LIT E5 01 2021 page 163 172...
Page 165: ...MK LIT E5 01 2021 page 165 172...
Page 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...
Page 173: ......