MK LIT (E5) 01/2021
page 23/172
3. Foreseeable Misuse
Other applications than those described in chap. 2 are not approved.
This expressly includes the following misuses (the list is not exhaustive), which pose risks despite the in-
herently safe construction and existing technical safety equipment:
•
Non-observance of Operating Manual
•
Non-observance of information and warnings on the chamber (e.g. control unit messages, safety iden-
tifiers, warning signals)
•
Installation, startup, operation, maintenance and repair by untrained, insufficiently qualified, or unau-
thorized personnel
•
Missed or delayed maintenance and testing
•
Non-observance of traces of wear and tear
•
Insertion of materials excluded or not permitted by this Operating Manual.
•
Non-compliance with the admissible parameters for processing the respective material.
•
Installation, testing, service or repair in the presence of solvents
•
Installation of replacement parts and use of accessories and operating resources not specified and au-
thorized by the manufacturer
•
Installation, startup, operation, maintenance or repair of the chamber in absence of operating instruc-
tions
•
Bypassing or changing protective systems, operation of the chamber without the designated protective
systems
•
Non-observance of messages regarding cleaning and disinfection of the chamber.
•
Spilling water or cleaning agent on the chamber, water penetrating into the chamber during operation,
cleaning or maintenance.
•
Cleaning activity while the chamber is turned on.
•
Operation of the chamber with a damaged housing or damaged power cord
•
Continued operation of the chamber during an obvious malfunction
•
Insertion of objects, particularly metallic objects, in louvers or other openings or slots on the chamber
•
Human error (e.g. insufficient experience, qualification, stress, exhaustion, laziness)
•
Non-observance of information and warnings on the chamber (in particular signals from the gas detec-
tion device)
•
Operation of the device and work on the chamber without personal protective equipment, if required
•
Insertion of defective accumulators or prototypes
•
Insertion of inadmissible battery types, e.g. lead batteries
•
Inserting modules (interconnected cells) or battery systems into the chamber during tests with EUCAR
hazard levels 4 to 6
•
Insertion of several cells with LIT MK 240 / of more than three cells with LIT MK 720 simultaneously into
the chamber without customer rinsing/inertization during tests with EUCAR Hazard Levels 4 to 6
•
Exceeding the permissible total energy or potentially released gas quantity
•
Failure to observe the minimum distances between the inserted batteries
Summary of Contents for 9020-0402
Page 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Page 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Page 162: ...MK LIT E5 01 2021 page 162 172...
Page 163: ...MK LIT E5 01 2021 page 163 172...
Page 165: ...MK LIT E5 01 2021 page 165 172...
Page 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...
Page 173: ......