MK LIT (E5) 01/2021
page 48/172
NOTICE
Danger of overheating due to lack of ventilation.
Damage to the chamber.
∅
Do NOT install the chamber in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
Observe the prescribed minimum distances when installing the chamber.
Do not install or operate the chamber in potentially explosive areas.
DANGER
Danger of explosion due to combustible dusts or explosive mixtures in the vicinity
of the chamber.
Serious injury or death from burns and / or explosion pressure.
∅
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.
Ambient conditions
•
Permissible ambient temperature range during operation: +18 °C /
64.4 °F
to +32 °C /
89.6 °F
. At ele-
vated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +25 °C /
77 °F
to which the specified technical data relate. For other ambient con-
ditions, deviations from the indicated data are possible.
•
Permissible ambient humidity: 70 % r.h. max., non-condensing.
When operating the chamber at temperature set-points below ambient temperature, high ambient humidity
may lead to condensation on the chamber.
•
Installation height: max. 2000 m /
6562 ft.
above sea level.
Minimum distances:
•
Distance between each chamber when placing several chambers of the same size side by side: 250
mm /
9.84 in
•
Wall distances: rear 300 mm /
11.81 in
, sides 200 mm /
7.87 in
•
Chambers with optional water cooling (without optional voltage and frequency changer): Wall distance
rear 100 mm /
3.94 in
.
•
Chambers with voltage and frequency changer: rear wall distance of the battery test chamber approx. 1
m /
3.28 ft
to set up the voltage and frequency changer
•
Spacing above the chamber: 100 mm /
3.94 in
The chambers are NOT intended for stacking.
NOTICE
Danger by stacking.
Damage to the chambers.
∅
Do NOT place the chambers on top of each other.
Summary of Contents for 9020-0402
Page 159: ...MK LIT E5 01 2021 page 159 172 30 7 Dimensions LIT MK 240 dimensions mm...
Page 160: ...MK LIT E5 01 2021 page 160 172 LIT MK 720 dimensions mm...
Page 162: ...MK LIT E5 01 2021 page 162 172...
Page 163: ...MK LIT E5 01 2021 page 163 172...
Page 165: ...MK LIT E5 01 2021 page 165 172...
Page 166: ...MK LIT E5 01 2021 page 166 172 31 3 Safety concept certificate from T V S d...
Page 173: ......