OPERON B 810 SURGICAL TABLE
SERVICE PROGRAM
70-000-0027 Rev.
1
6-21
No.
1
Error
Reaction Description
Solution
17
Motor “current
feedback” fault
Continue Amperage exceeds expected current draw
from Motor (Expected maximum is 14.3
amps)
Motor short circuit, Motor disconnected,
broken wiring
•
Locate and eliminate
short circuit.
•
Connect motor.
•
Replace broken
wire(s).
19
12 V fault
Continue Power Supply 12 volt fault Voltage
Regulator possibly damaged
•
Check battery and the
charging circuit
connections.
•
Refer to Figure 5-2
and check circuits to
and form J2 on the
CPU board. Replace
faulty harnesses or
components.
20 5
V
fault
Continue Power Supply 5 volt fault, Voltage Regulator
possibly damaged
Refer to Figure 5-2 and
check the sensor circuits.
Replace defective
harnesses or
components.
21
Ground or ADC
fault
Continue Ground wiring possibly broken, ADC of the
Micro controller defective
•
Check ground wiring
continuity.
•
Replace CPU.
22 Sensor-1
Fault
(Tilt)
Continue Sensor disconnected, wiring broken, short
circuit
•
Connect sensor.
•
Identify and replace
broken wiring.
•
Locate and remove
cause of short circuit.
23
Sensor-2 Fault
(Trend)
Continue Sensor disconnected, wiring broken, short
circuit
•
Connect sensor.
•
Identify and replace
broken wiring.
•
Locate and remove
cause of short circuit.
24
Sensor-3 Fault
(Back)
Continue Sensor disconnected, wiring broken, short
circuit
•
Connect sensor.
•
Identify and replace
broken wiring.
•
Locate and remove
cause of short circuit.
25
Sensor-4 Fault
(Legs)
Continue Sensor disconnected, wiring broken, short
circuit
•
Connect sensor.
•
Identify and replace
broken wiring.
•
Locate and remove
cause of short circuit.
26
Sensor-5 Fault
(Lock)
Continue Sensor disconnected, wiring broken, short
circuit
•
Connect sensor.
•
Identify and replace
broken wiring.
•
Locate and remove
cause of short circuit.
Summary of Contents for OPERON B 810
Page 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Page 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Page 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Page 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Page 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Page 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Page 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Page 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Page 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...