OPERON B 810 SURGICAL TABLE
OPERATING INSTRUCTIONS
Rev. 1
70-000-0027
2-8
REMOVABLE LEG SECTION
WARNING
WARNING
— Failure to securely
lock the leg section into place may
result in patient injury. Always check
the retention of the leg section upon
installation.
Install the leg section as follows:
1.
Using the pendant, activate LEG UP until the leg
section pivots are horizontal.
2.
To align the leg section for installation, place the
leg section posts into the pivots at a –45 degree
angle.
FIGURE 2-8
ALIGNING THE LEG
SECTION FOR INSTALLATION
3.
Once the leg section is aligned, lift to a horizontal
position and insert the leg section into the pivots.
4.
Slide the leg section towards the seat section until
it comes to a stop and the locks snap into place
(FIGURE 2-9 and FIGURE 2-10).
Leg Section
Seat Section
Leg Section Pivots
FIGURE 2-9
LEG SECTION
Leg
Section
Latches
Seat
Section
FIGURE 2-10
ASSEMBLED LEG
SECTION
5.
Attempt to pull the leg section away from the
table to make sure that it is retained. If not, push
evenly until the locks engage.
To remove the leg section:
1.
Bring the leg section to a level position, (i.e.,
parallel to the floor).
2.
Depress the latches to release the locking
mechanism and pull back evenly.
FOOTSWITCH (OPTIONAL)
CAUTION
CAUTION
— The Footswitch’s
electrical connections are resistant to
fluid intrusion, but if fluid drainage
is anticipated, the footswitch should
be protected by wrapping it with a
clear disposable footswitch bag.
This will reduce the need to disinfect
the footswitch.
NOTE
NOTE
— to operate the footswitch
control, the floor feet must be locked.
NOTE
NOTE
— It is not necessary to press
the primary hand pendant ON/OFF
switch when operating the footswitch
control. The primary hand pendant
can not be operated simultaneously
with the footswitch.
The footswitch plugs into the left connector on the
receptacle box, which is located on the left side of the
table. Align the red dot on the footswitch connector
with the red dot on the mating connector on the left
side of the connector box. (See FIGURE 2-11.)
Fully Inserted
Pivots Level
Partially Inserted
Summary of Contents for OPERON B 810
Page 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Page 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Page 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Page 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Page 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Page 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Page 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Page 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Page 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...