OPERON B 810 SURGICAL TABLE
SERVICE PROGRAM
70-000-0027 Rev.
1
6-13
ACCU Tab
The ACCU (accumulator) tab sets the battery power
indicator display for the hand pendant (Figure 6-15
and TABLE 6-2). These are the voltage readings that
cause the LED display to change from one level to
the next higher level as the battery accumulates a
charge. The voltages should be set toward the upper
end of the range to ensure that the table has adequate
time to charge before being unplugged from the AC
mains.
CAUTION
CAUTION —
The Service Program will
allow you to set ACCU parameters
outside the ranges shown in TABLE 6-2.
This could lead to false readings on the
(physical and graphic) pendant LEDs,
including a failure to operate. Always
double check your settings before
clicking on “Save Parameters”.
To set the value of the parameter, use your cursor to
highlight the number you wish to change. Type in
the new value to replace the old value.
The Status window on the tab shows the current
battery charge level in VDC. A bar graph under the
voltage value shows the charge as a percentage of
total capacity. These are display values based on
CPU readings; they cannot be changed in the field.
The shut down time slide sets the sleep time delay:
how long after the last use until the hand pendant
enters “sleep” mode. In sleep mode, the back
lighting goes out and all LEDs turn off. The longer
the table remains in “sleep” mode, the longer the
battery charge will last.
User preference should be the primary consideration
on how long the sleep time delay should be.
Generally, procedures which require more frequent
table position changes should have longer sleep time
delays. Procedures that require infrequent position
changes can have shorter sleep time delays.
To set the sleep time delay, click and hold the left
mouse button as you drag the pointer to the desired
setting. Time increments are in full minutes.
After setting all of the parameters to the desired
levels, click on the “Save Values” button to record
them in the table’s CPU. (Remember to save the
settings to both CPUs.)
FIGURE 6-15.
ACCU (BATTERY) DISPLAY SETTING FORM
TABLE 6-2.
ACCU (BATTERY) SETTING PARAMETERS
Name Default
Limit 10% (LED changes from Steady Red to Flashing Red)
21.4V
Limit 30% (LED changes from Flashing Red to Yellow)
22.7V
Limit 90% (LED changes from Yellow to Green)
23.5V
Sleep timer
120 minutes
Summary of Contents for OPERON B 810
Page 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Page 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Page 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Page 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Page 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Page 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Page 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Page 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Page 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...