OPERON B 810 SURGICAL TABLE
SERVICE PROGRAM
Rev. 1
70-000-0027
6-26
TABLE 6-4.
TROUBLESHOOTING GUIDE
Problem Action
If battery indicator is green or yellow, no action is required. The batteries
are charged.
If the battery indicator is red (flashing), use the manual foot pump to
provide hydraulic power.
Loss of main AC electrical power,
indicated by the charging indicator
being OFF.
If the battery indicator is red (steady), use the auxiliary pendant to select
functions and the manual foot pump to provide hydraulic power.
Main AC power is present, but the
charge indictor does not illuminate.
Check mains AC fuses (see Figure 6-20). Replace if blown. If this does
not correct the problem, an internal table failure has occurred that requires
repair. Contact Berchtold Service for assistance.
The main DC power switch on the base must be in the ON position (up).
If it is down (OFF), push on the top of the switch to turn it to the ON
position (see Figure 6-20).
Primary hand pendant will not turn
on.
The Battery fuse is blown.
•
Check the 2nd from the top fuse in the drawer (Figure 6-20).
•
Replace if blown.
•
Take a current reading to check for an over current condition.
•
Isolate and replace failed parts.
•
Plug table into AC power supply for use if no fuse is available.
Check the status of the Lock LED. The table must be locked to the floor
to enable the other functions to operate. If the table is unlocked, press
LOCK to lock the table to the floor.
Primary hand pendant illuminates
when turned on, but will not function.
Check the battery status LED. If it shows a steady red light, the CPU has
commanded an automatic shut down and disabled the hand pendant. If
possible, connect the power cord to an AC power source. If no AC power
source is available, use the auxiliary pendant (and foot pump, if necessary)
to complete the procedure.
Primary hand pendant failure.
Check all pendant connections. If all are satisfactory, replace pendant.
Use the auxiliary hand pendant (and the foot pump if necessary) to
complete the procedure. Use the service program to check the function
Head section or leg extension cannot
be removed.
Ensure that the levers on each end of the head or leg section are fully
depressed. Pull the head section or leg extension out by applying an even
force on each side of the section.
Service indicator illuminates.
An illuminated service indictor can cause certain functions, such as RTL,
CHAIR, AND FLEX to be disabled. Other functions may work normally.
Use the Service Program to determine the cause, then repair or replace the
component at fault.
The motor may be overheated. Let the table sit for 15 minutes before
attempting another function. The auxiliary pendant and foot pump can be
used to complete the procedure.
Check fuse F4 in the left power center (Figure 6-20). Replace if blown.
Check for over current on pump. Replace pump if defective.
The primary hand pendant functions
normally, but the pump motor will
not operate.
The pump or the CPU may have failed. If pump is good, replace CPU.
Summary of Contents for OPERON B 810
Page 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Page 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Page 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Page 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Page 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Page 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Page 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Page 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Page 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...