OPERON B 810 SURGICAL TABLE
GENERAL INFORMATION
70-000-0027
Rev.
1
1 - 5
Primary Hand Pendant Functions
WARNING
WARNING
— The primary hand pendant
is configured for normal patient
orientation. If the patient is reversed, the
back function must be used to raise or
lower the patient's legs and the leg
function must be used to raise or lower
the patient's back. Trendelenburg, tilt and
slide functions are also reversed if the
patient is in reverse orientation.
1.
BACK
v
— Raises and lowers the back section.
2.
TABLE
v
— Raises and lowers the entire
tabletop.
3.
LEG
v
— Raises and lowers the leg section.
4.
TREND/REV TREND
(Trendelenburg) — Tilts
the entire tabletop about the transverse axis.
5.
SLIDE FOOT/SLIDE HEAD
— Slides the
table top toward and away from the head.
6.
TILT LEFT/TILT RIGHT
— Tilts the entire
tabletop about the head to foot axis.
7.
UP KIDNEY/DOWN KIDNEY
— Raises and
lowers the kidney elevator.
8.
FLEX
— Combines back down with reverse
Trendelenburg to form an inverted V (jackknife
position).
9.
CHAIR
– Combines back up, Trendelenburg,
and leg down functions to form a chair-like
support. The three functions operate in a
repeating sequence until the back section reaches
80° relative to the seat section and the leg
sections reach 105° degrees relative to the seat
section.
10.
LEVEL
— Moves the table to a horizontal
position that is within 1 degree of level. This
function should be used prior to the Flex and
Chair Functions.
11.
ORIENTATION NORMAL/REVERSE
— In
REVERSE mode, the buttons function in reverse.
CHAIR is disabled in Reverse Orientation. Table
up/down, floor lock/unlock, and kidney up/down
are not reversed.
12.
FLOOR
UNLOCK
/
LOCK
— Raises
(unlocks) or lowers (locks) the floor feet.
NOTE
NOTE: — The power switch on the base
of the table must be in the ON position
before the hand pendant will operate.
13.
ON/OFF
— Turns the primary hand pendant on
and off. The pendant will be backlit when
activated.
NOTE
NOTE
— If AC power is not present, the
auto-shutdown feature will turn the
control pendant off after about 120
minutes of non-use. To reactivate the
pendant, press any function button. If
AC is present, the auto-shutdown feature
is disabled, and the hand pendant
remains on until it is turned off.
1
2
5
4
3
6
7
8
9
10
11
12
13
FIGURE 1-6
PRIMARY HAND PENDANT
FUNCTIONS
Summary of Contents for OPERON B 810
Page 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Page 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Page 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Page 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Page 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Page 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Page 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Page 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Page 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...