OPERON B 810 SURGICAL TABLE
PRINCIPLES OF OPERATION
Rev. 1
70-000-0027
4-22
The charge indicator LED illuminates green when
AC is applied to the table and the batteries are
charging. The indicator should always be illuminated
if AC is applied.
Charging LED
Ground Stud
AC Fuse Drawer AC Fuses
Access Cover
For AC Mains
Connection
FIGURE 4-36.
RIGHT POWER CENTER
CONTROL PENDANTS
Primary Hand Pendant
WARNING
WARNING – The table has two hand
pendants that can control its
movements. Both can be active at the
same time. If different functions are
pressed on the different hand
pendants, both functions will activate.
This could lead to unexpected or
unintended movement of the table
sections and could result in injury to
patients or personnel or in damage to
the equipment. To avoid this, keep the
auxiliary hand pendant securely
stowed in its drawer during normal
usage. When the auxiliary hand
pendant is deployed during an
emergency, unplug the primary hand
pendant to prevent its use.
Commands are entered on the primary hand pendant
through the keypad.
The primary hand pendant sends commands to the
CPU via connection J6 on the CPU board.
An illustration of the primary hand pendant along
with function and LED display descriptions is
provided in the General Information section of this
manual.
If the primary hand pendant is not communicating
with the CPU, the SERVICE (Service) LED will
blink.
If no AC is connected to the table, both the primary
hand pendant and the table will automatically go into
sleep mode 2 hours after the last function was
pressed. The table may be powered up again by
pressing any button on the primary hand pendant.
Pressing ON while a function is running will stop the
function. Pressing any other function button while a
function is running will stop the articulation. To start
the articulation, release the function button and press
it again.
Pressing OFF while a function is running will stop
the function and turn the table off.
If AC is connected to the table, the auto-shutdown
feature is disabled (i.e., the hand pendant and table
operations will not automatically turn off after 2
hours of non activity).
The auxiliary hand pendant drawer does not have an
affect on the primary hand pendant’s power.
Removing the auxiliary pendant activates the
auxiliary pendant, but it does not deactivate the
primary hand pendant. Thus, if a function is pressed
on the primary pendant while another function is
pressed on the auxiliary pendant, both valves will
open and pump pressure will activate both
movements.
There are no field serviceable parts in the primary
hand pendant. It must be replaced as a complete
assembly.
Summary of Contents for OPERON B 810
Page 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Page 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Page 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Page 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Page 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Page 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Page 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Page 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Page 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...