charging will occur whether the ON/OFF switch on
the base is in the ON or OFF position. The table can
remain plugged into AC indefinitely without
damaging the batteries.
HAND PENDANT OPERATION
The table may be activated by pressing the hand
pendant ON button or by pressing any of the hand
pendant function buttons.
If you activate the table by pressing the ON button,
the hand pendant LEDs for the SERVICE,
LOCK/UNLOCK, and all Battery levels will
illuminate briefly. This is a self-test function.
If there is a problem with the hand pendant, the
connection to the table, or any of the table functions,
the SERVICE LED will remain on.
If there are no problems, all LEDs except the current
floor lock and battery status displays will turn off,
and the function buttons will become back lit to
indicate that the table is in a “controls active” status
(FIGURE 2-6).
Lock or Unlock
Status LED stays lit
Battery
Status LED stays lit
Function
Indicator
Back
Lighting
Glows
SERVICE Led
Lights then goes out
FIGURE 2-6
HAND PENDANT POWER-ON
INDICATORS
To conserve battery power, the hand pendant will
automatically go into SLEEP MODE (backlighting
will turn off) after two hours of non-use. This period
can be changed through the table’s Service Software
Program, which is described in the “Service
Program” section of this manual.
If you activate the hand pendant from the SLEEP
MODE by pressing any of the function buttons (other
than the ON button), the self-test will be performed
after the button is released. To actuate the function
associated with the button it must be pressed a second
time after “awakening” the hand pendant. If the
floor legs are NOT locked, the back lighting will
come on and the hand pendant will go into “controls
active” status, the feet will automatically descend and
lock.
To operate any hand pendant function, press and hold
the button for the action desired. If you press the floor
lock or unlock function, you do not have to hold the
button until the function is complete. Once started, the
floor locking or unlocking action will be completed
automatically.
Each articulation has a software controlled limit. You
can stop any motion by releasing the function button
or you can hold the button until it reaches the limit set
for that function, and the movement will stop
automatically.
1.
Press the ON button on the control pendant to
activate the control pendant
2.
Press FLOOR LOCK to lower the floor feet. The
green LOCK indicator on the control pendant will
light when the floor feet are locked.
3.
Select the desired function by pressing and
holding the appropriate function switch.
4.
When the desired position has been reached,
release the function button to stop movement.
5.
After the patient has been transferred from the
table, press FLOOR UNLOCK to raise the floor
feet and allow table transport. The yellow
UNLOCK indicator on the control pendant will
light when the floor feet are unlocked.
If the table functions do not operate, check the
following:
1.
Is the ON/OFF switch on the base of the table in
the OFF position?
•
Turn the switch to the ON position.
2.
Is the yellow UNLOCK indicator lit?
•
Lock the floor feet.
3.
Is the battery status LED red and flashing or
steady?
•
Use the manual foot pump or plug in the
table. Plugging in the table will allow normal
use and also recharge the batteries.
•
Use the auxiliary pendant and foot pump to
complete the operation.
4.
Is the red SERVICE LED lit?
•
A lit SERVICE LED indicates that service is
required, but may not disable a table function
or functions, depending on the failure. If a
failure occurs that can allow an unsafe
operation, the specific function, such as
return to level, will be disabled but the other
functions will remain enabled. Only a
catastrophic failure of the main processor will
light the SERVICE LED and disable all table
functions. Always contact Berchtold Service
if the SERVICE LED is lit.
Summary of Contents for OPERON B 810
Page 1: ...OPERON B 810 SERVICE PARTS MANUAL ...
Page 71: ...OPERON B 810 SURGICAL TABLE SERVICE PROGRAM 5 8 Rev 1 70 000 0027 BASE WIRING DIAGRAM ...
Page 104: ...OPERON B 810 SURGICAL TABLE PARTS 7 4 Rev 1 70 000 0027 THIS PAGE INTENTIONALLY BLANK ...
Page 109: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 5 ...
Page 110: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 6 ...
Page 116: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 12 BERCHTOLD Corp BERCHTOLD Corp ...
Page 117: ...OPERON B 810 SURGICAL TABLE PARTS 70 000 0027 Rev 1 8 13 B E R C H T O L D C o r p ...
Page 118: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 14 BERCHTOLD Corp BERCHTOLD Corp ...
Page 120: ...OPERON B 810 SURGICAL TABLE PARTS Rev 1 70 000 0027 8 16 BERCHTOLD Corp BERCHTOLD Corp ...