Operation and setting
45
Connecting the test probe during zero balancing
1. A DUT has a detachable power supply cord. If only the resistance of the PE conductor is to be
measured, make the connection as follows:
How to perform zero balancing for measurement # 2:
– Insert the adapter into bush “C”.
– Contact the adapter's PE contact with the PE contact on test socket “E”.
– Start zero balancing by pressing “F1”.
2. A DUT has a detachable power supply cord. If only the resistance of the DUT is to be measured,
make the connection as follows:
How to perform zero balancing for measurement # 1:
– Insert the measuring cable (with test terminal if applicable) into bush “C”.
– Insert the passive test probe into bush “B” or active test probe into the colour-coded bushes
“B” and “A”.
– Contact the test probe with the end of the measuring cable (with test terminal if applicable).
– Start zero balancing by pressing “F2”.