Operation and setting
40
4.5.6
Setup
[-> Main menu -> F5: System administration -> F5: Setup]
The following settings can be made in the “Setup” menu:
4.5.6.1
Language
[-> Main menu -> F5: System administration -> F5: Setup -> F1: Language]
All menus will appear in the selected language. “Austrian” is used to adapt menu and
protocol texts to reflect standards applicable in Austria (e.g. ÖVE-E 8751-1).
4.5.6.2
Select nominal voltage
[-> Main menu -> F5: System administration -> F5: Setup -> F2: Select nominal voltage]
The UNIMET® 1100ST supports a voltage range between AC 90 and 264 V . In order to
always obtain comparable measured values even if the system voltage is fluctuating, various
standards require measured values to be converted to nominal system voltage or even to
106% or 110% of the nominal system voltage.
For this reason, you should set the nominal system voltage. The factory setting is 230 V.
UNIMET® 1100ST automatically performs the required conversion of the measured values.
,
Select language. Press Enter to confirm your selection.
Exit function without making changes.
This setting does not affect the internal power supply of the test device or the
voltage applied at the test socket. UNIMET® 1100ST provides the DUT with the
same voltage it is supplied with itself.
<<