Testing and measuring
52
5.2 Classification
Select the applicable test standard from the main menu. Answer the questions that appear
on the screen. The test system will identify the required tests and limiting values to be
observed. Save the classification under the device’s type name in the UNIMET® 1100ST type
catalogue.
Example 1:
Classification of a medical electrical device (e.g. combined ultrasound/stimulation current
unit) to DIN VDE 0751-1:2001-10. This is a Class I equipment with two patient connections.
In the main menu, select “F3: Classification”.
5.2.1
Test standard
Select the test standard in accordance with which the device is to be tested.
Example: “F2: DIN VDE 0751-1:2001-10”.
5.2.2
Protection class
Select the DUT protection class in accordance with which the device is to be tested.
Example: “F1: Protection class I (Class I equipment)”
(Note: F5: Hospital bed test = Abbreviated classification for hospital beds)