UniCel DxC 600i In-Lab Training Manual A24574-AD
Page 8
Professional Development Department
October 2010
Cartridge Chemistry (CC) System
The Cartridge Chemistry System provides random access analysis of
analytes chosen by the user. Reagents are delivered from reagent
cartridges stored in a refrigerated compartment that houses up to 59
cartridges. The reactions occur in permanent glass cuvettes that are
housed in a reaction carousel. Reagents are delivered to the reaction
carousel by reagent probes working in conjunction with a reagent
syringe. Sample is added by a CC sample probe and syringe. Mixers
are responsible for mixing the reagent and sample. Reactions are
read by a photometer assembly. A cuvette wash station cleans the
cuvettes for reuse.
Hydropneumatic System
The Hydropneumatic System is responsible for providing vacuum,
compressed air, diluted wash solution, and deionized water.
Operation and Control Components
Basic operating functions are controlled and reviewed by utilizing the
DxC 600i monitor, mouse, touch screen, keyboard, and push-button
controls. These provide the interface between the operator and the
DxC.
Additional detailed information about CTA components, DxC 600
components, and Access components can be found in the following
References:
CTA: UniCel DxC 600i Synchron Access Clinical Systems
(Online) Instructions For Use Manual
DxC 600: UniCel DxC Synchron Clinical Systems (Online)
Instructions For Use Manual
Access: Access Online HELP System