UniCel DxC 600i In-Lab Training Manual A24574-AD
Page 58
Professional Development Department
October 2010
1. Select
Samples
from the DxC 600i monitor
2. Enter the
Sample ID
3. Enter
rack
and
position
if bar code is
unavailable or unreadable
4. Select
Chemistries
if programming is not
available
If needed, enter additional information:
select the
STAT
check box
enter a
Sample Type
Enter a
Sample Comment
Enter demographic information using
Demog [F2]
5. Select
Next [F10]
6. Place the sample in the rack and position
identified above (push the tube all the way
down)
Place the bar code labeled sample container in the
appropriate rack. Ensure that the entire length of
the bar code label is visible through
the rack slot. Push the tube all the way down.
1. Open the CTA cover and place the rack in the
DxC 600i load area with the bar code facing to
the right
2. Press the
<RUN>
button on the CTA
Does
sample have LIS
programming and a
readable bar code
label?
No
Yes
Remove visible blood from the top of capped
tubes using a cotton tipped swab
moistened with DI water
No
*Adequate
Sample Volume?
Yes
Use the “Low Volume Sample Processing”
flowchart
*Adequate
sample
volume is
determined
by Primary
Tube Sample
Template
Sample Processing