UniCel DxC 600i In-Lab Training Manual A24574-AD
Page 70
Professional Development Department
October 2010
Access Commands
The Access system has commands that can be used to Initialize
(home), Pause, Shutdown, or set the CTA off-line. These commands
are initiated at the Access monitor.
Initialize:
•
Use to move mechanical assemblies to their “home” position
•
Use to return the system to
Ready
from the
Not Ready/Stopped
state
Pause:
•
Prevents new or additional tests from initiating at the Access
instrument
•
Any tests already in process are completed
•
The Access cannot be paused during a maintenance routine
•
The system may enter the
Paused
mode if a rack is
loaded/unloaded or a reagent is loaded/unloaded
Shutdown:
•
Use to reboot or safely shutdown the system
•
Files are safely closed to prevent data corruption
•
Failure to follow recommended procedure may result in data
corruption
Place the CTA unit off-line:
•
Removes the Access analyzer from control of the CTA
•
Use before conducting diagnostic or troubleshooting activities on
the Access analyzer
•
Use before homing Access components