UniCel DxC 600i In-Lab Training Manual A24574-AD
Page 31
Professional Development Department
October 2010
From the Access monitor,
select the <
Substrate>
button
Scan the bar code on the new, equilibrated substrate
bottle using the Access hand held bar code reader
Remove the cap from the new substrate bottle
Discard empty substrate bottle
Unscrew the substrate supply cap from the bottle in use
and immediately place the inlet tubing and cap
onto the new bottle
Place the new bottle back in the fluids tray
Select
Change Substrate [F5]*
Select
Done [F1]
Note: Access
must be in
READY mode
to change
substrate
Place a new bottle of substrate in the fluids tray to
equilibrate to room temperature
A prompt will appear asking if you want
to prime the substrate
Select
Yes [F1]
* The change
substrate
window is
displayed
Change Substrate
Summary of Contents for DxC 600i