Chapter 5
-DxC 600i Instrument
Commands
DxC 600i Instrument Commands
The DxC 600i instrument commands can be used to control functions
of DxC and CTA instrument components. Instrument commands
include:
Home:
•
Moves mechanical assemblies to their “home” positions and
primes the system
•
Recovers from a motion error
•
Returns the system to
Standby
from the
Stopped
state
Pause:
•
Prevents initiation of new or additional tests at the DxC or CTA
•
Tests in process are completed
•
Use to return the system to standby as soon as possible
Shutdown:
•
Use to reboot or safely shutdown the system
•
Files are safely closed to prevent data corruption (Failure to follow
recommended procedure may result in data corruption)
Enable/Disable Modules:
•
Use to enable/disable an individual MC cup module, the MC ISE,
the entire MC side, the CC Photometer, the LPIA module, the CC
lower reagent carousel, or the entire CC side of the system
•
All programming for disabled modules is aborted. When a module
is disabled the system will not run, prime, perform diagnostics, or
maintenance procedures using that module
•
The system may disable modules automatically after certain
unrecoverable error conditions
•
When a module is enabled, it will automatically prime
•
When all cup modules are disabled, the ISE module is
automatically disabled
Unload All:
•
Use to unload racks from the DxC/CTA sample carousel
•
Use when access to a sample(s) is required
•
Use when the screen displays racks on the Main Screen but the
racks are not on the sample carousel following a system error
Other instrument commands include:
<3> Stop Print
,
<5> Pause
Waste B
, and
<6> Resume Waste B
. Information about these
commands may be found in the UniCel DxC Synchron Clinical
Systems Instructions For Use Manual.