UniCel DxC 600i In-Lab Training Manual A24574-AD
Page 42
Professional Development Department
October 2010
Unfasten the cap and remove the old Auto-Gloss
bottle
Use the DxC hand held bar code reader to scan the
bar code on the Auto-Gloss bottle when the
Auto-Gloss data screen appears.
Verify the Auto-Gloss data is displayed
Select
Rgts/Cal
from the DxC 600i monitor
Open the CTA door. The Auto-Gloss bottle is
located on the bottom shelf
Select the
Supplies
tab
Select the
CTA Auto-Gloss
Select
Load [F1]
Place the new Auto-Gloss bottle on the system, and
tighten the cap
Select the
Done [F1]
Change the CTA Auto-Gloss