Shutdown the DxC 600i System
The DxC 600i system may require power down when:
•
Parts replacement procedures specify power to be turned off
•
The lab power goes off
•
The system is moved to a new location
•
Recommended by a BCI specialist for troubleshooting
purposes
Avoid system errors by following these recommendations:
•
The CTA unit must be powered down whenever the DxC is
powered down
•
The CTA may be powered down/up without rebooting the
DxC analyzer if there are no racks on the CTA sample wheel
or no aliquotted AVs at the AV queue
Note: The Access system is powered down separately at the
Access analyzer.
Use the flowcharts on the following pages when powering down
the DxC 600i system.