UniCel DxC 600i In-Lab Training Manual A24574-AD
Page 41
Professional Development Department
October 2010
No
Yes
Gently invert a new bottle of wash buffer
3-4 times
Remove the cap and inner seal from the
new bottle
Remove empty bottle and remove the dispense
cap assembly from the empty bottle
Attach dispense cap assembly to new bottle
Turn new bottle upside down and place into
reservoir receptacle
Lot number tracking
required?*
Select
Rgts/Cal
from the
DxC 600i monitor
Select the
Supplies
tab on the screen
Procedure
Complete
Type the lot number into the CTA Wash
Buffer field located on the CTA Supplies
section of the screen
Open the CTA door and slide out the metal
tray to gain access to the wash buffer
Close the CTA door
*Lot number tracking
requirement defined by
laboratory policy
Change the CTA Wash Buffer