C-1
Z MED Medical Air System
4107 9022 67.00
C.1 Safety Precautions
C.1.1 General Precautions
1. The operator must employ safe working
practices and observe all related local work
safety requirements and regulations.
2. If any of the following statements does not
comply with local legislation, the stricter of
the two shall apply.
3. Installation, operation, maintenance, repair
work must only be performed by authorized,
trained, specialized personnel.
4. Before carrying out any maintenance, repair
work adjustments or any other non-routine
checks, stop the device. In addition, the
power isolating switch must be opened and
locked.
5. Use only correct tools for maintenance and
repair work.
6. Use only genuine
BeaconMedæs
spare parts.
Contact
BeaconMedæs
Technical Services
department at 1-888-4MEDGAS (1-888-463-
3427) for technical support or to order spare
parts.
C.2 Compressor Control Panel
Parts and Functions (Mk5)
1. Display:
Shows the compressor operating
condition and a number of icons to navigate
through the menu.
2. Pictograph:
Automatic operation
3. Pictograph:
General alarm
4. General Alarm LED:
Flashes if a shutdown
warning condition exists.
5. Pictograph:
Service
6. Service LED:
Lights up if service is needed.
7. Automatic Operation LED:
Indicates that
the regulator is automatically controlling the
compressor.
8. Voltage On LED:
Indicates that the voltage is
switched on.
9. Pictograph:
Voltage on
10. Enter Key:
Key to select the parameter
indicated by the horizontal arrow. Only the
parameters followed by an arrow pointing to
the right can be modified.
11. Escape Key:
To go to previous screen or to
end the current action
12. Scroll Keys:
Keys to scroll through the menu
13. Stop Button:
Button to stop the compressor.
LED (7) goes out.
14. Start Button:
Button to start the
compressor. LED (7) lights up indicating that
the Elektronikon regulator is operative.
Appendix C - Compressor Controller (Mk5)
WARNING!
All responsibility for any damage or
injury resulting from neglecting these
precautions, or non-observance of the
normal caution and care required for
installation, maintenance and repair,
even if not expressly stated, will be
disclaimed by the manufacturer.
FIG C.1 Compressor Control Panel
C.3 Icons Used (Mk5 Controller)