Z MED Medical Air System
4107 9022 67.00
Problem
Possible Causes
Solution
Compressor cycles
too often
System undersized
Incorrect pressure setting
Faulty pressure transducer
System piping leaks
Safety valve leaks
Both dryers on line
Water in air receiver
Contact
BeaconMedæs
Adjust pressure band parameters
Replace transducer
Repair leaks
Repair or replace valves
Valve off one dryer
Drain air receiver
Compressor won’t
shut off
Faulty pressure transducer
System piping leaks
Adjust or replace
Repair leaks
High air temperature
Bad room ventilation
Air Filter Clogged
Insufficient cooling water (ZR)
Formation of scale/dirt
Improve room ventilation
Replace filter
Check water temperature, increase
water flow
Contact
BeaconMedæs
Dryer not Cycling
Main power disconnected
Dryer power switch off
Power failure
Main fuse blown
Fuse blown in control circuit
Solenoid valve failure
Smart Relay failure
Loose or faulty connection
Turn on main power
Turn on power switch
Restore power
Replace fuse
Replace fuse
Check and replace or repair solenoid
valve
Check and replace if defective
Check & tighten all wire connections