Z MED Medical Air System
4107 9022 67.00
Safety Precautions
Pressurized air from the system may cause
personnel injury or property damage if the unit
is improperly operated or maintained.
Operator should have carefully read and become
familiar with the contents of this manual before
installing, wiring, starting, operating, adjusting
and maintaining the system.
Operator is expected to use common sense
safety precautions, good workmanship practices
and follow any related local safety precautions.
In addition:
• Before starting any installation or
maintenance procedures, disconnect all
power to the package.
• All electrical procedures must be in
compliance with all national, state, and local
codes and requirements.
•
A certified electrician should connect all
wiring.
• Refer to the electrical wiring diagram
provided with the unit before starting any
installation or maintenance work.
• Release all pressure from the package
before removing, loosening, or servicing any
covers, guards, fittings, connections, or other
devices.
• Notify appropriate hospital personnel if
repairs or maintenance will affect available
compressed air levels.
• Air inlet must be placed in an area free of
toxic or hazardous contaminants. It must be
kept away from ETO exhaust vents, vacuum
exhaust vents, areas close to automotive
exhausts, etc., in accordance with NFPA 99.
• Prior to using the Desiccant Dryer System, the
medical facility must have a Certifier perform
all installation tests as specified in NFPA 99.
The medical facility is also responsible for
ensuring that the Medical Air meets the
minimum requirements as specified in NFPA
99.
• This is a high speed, rotating piece of
machinery. Do not attempt to service any
part while machine is in operation.
• To prevent automatic starting, disconnect
all electrical power before performing any
maintenance.
• Make sure that all loose articles, packing
material, and tools are clear of the package.
• Never operate a compressor with its isolation
(shutoff ) valve closed or without its relief
valve in place. Damage to the compressor
may occur.
• Check all safety devices periodically for
proper operation.
• Electrical service must be the same as
specified on the control panel nameplate or
damage to the equipment may occur.
• Vibration during shipment can loosen
electrical terminals, fuse inserts, and
mechanical connections. Tighten as
necessary.
G.3 Operation
G.4 Reading and Measurements
G.5 Oil Concentration Determination
G.6 Oil Indicator Replacement Cartridge