F-3
Z MED Medical Air System
4107 9022 67.00
Appendix F - CO Transmitter
5. Locate the dark blue pot (marked “C”)
and adjust the pot until the meter reads
0.051 VDC. Turning the adjusting screw
counterclockwise will reduce the voltage
output. Turning it clockwise will raise the
voltage output.
6. Repeat the zeroing procedure and
calibration procedure once more to ensure
the transmitter is set properly.
7. Reinstall the cover on the transmitter.
8. Reconnect the inlet air line from the dryer.
FIG F.9.1 CO Transmitter Assembly