F-4
Z MED Medical Air System
4107 9022 67.00
Appendix F - CO Transmitter
Description
Part Number
Qty Required
CO Sensor
4107 6530 69
1
Calibration Kit
*
4107 4004 63
1
20 ppm Carbon Monoxide
4107 6525 14
1
Impurity Free Air (Zero Gas)
4107 6525 13
1
F.10 Troubleshooting
F.11 Accessories and Replacement
Parts
WARNING!
Before removing the CO sensor, verify
that line pressure has been valved off or
reduced to atmospheric pressure.
Before servicing the CO sensor, do the follow-
ing:
1.
Unplug sensor(s) or turn off monitor.
2. Depressurize the CO sensor.
Note:
Remote alarms will be activated.
Problem
Possible Causes
Solution
CO Alarm - 10 ppm or above
Inlet air to the medical air system
is contaminated
Move air inlet location or remove
source of CO contamination
Monitor will not calibrate
Sensor depleted
Replace sensor
CO Alarm - CO reading erratic
Incoming power feed to control
panel adjacent to CO signal wires
Separate incoming power feed from
CO signal wires
(
*
) Kit includes calibration connector, 20 ppm test gas, and 0 ppm test gas in carrying case.