B-5
Z MED Medical Air System
4107 9022 67.00
Appendix B - Compressor Controller (Mk4)
B.5.2 Configuration Listing
Name
Description
Controller
Default Value
Beacon
Medaes Value
'Power Recovery Time'
Maximum allowed power down time to accept
the Automatic Restart After Voltage Failure
(ARAVF)
3600 seconds
Infinite*
'C.C.M. '
ES6 Control Mode (Local, Remote, LAN)
Local
LAN
'Pressure Band Selec'
Allows manual selection between pressure
band 1 or 2
Press Band 1
Press Band 1
'Auto Restart'
Allows the ES-6 to restart after a power failure Not Activated
Infinite
‘Password’
This feature allows the parameters to be pro-
tected by a user defined password
Not Activated
Activated**
‘Language’
English
English
'Time'
E.S.T.
E.S.T.
‘Date’
Current
Current
‘Date Format’
DD/MM/YY
MM/DD/YY
'Pressure Units'
Bar
psi
‘Temperature Unit’
°C
°C
‘Vibration Units’
microns
microns
‘Level Units’
mm
mm
‘Flow Units’
LPS
CFM
‘Volume Units’
Liter
Liter
* Controller display will show “oo! Sec” when “Auto Restart” parameter is set to infinite
** Requires password to change parameters and configuration (password = 4735)