A-9
Z MED Medical Air System
4107 9022 67.00
Name
Description
Controller
Default Value
Beacon
Medaes Value
‘Pressure Band 1 Low’
Low pressure level of Pressure band 1
7 bar
90 psi
‘Pressure Band 1 High
High pressure level of Pressure band 1
8 bar
105 psi
‘Pressure Band 2 Low’
Low pressure level of Pressure band 2
7 bar
90 psi
‘Pressure Band 2 High’
High pressure level of Pressure band 2
8 bar
105 psi
‘Number of Comp.’
Number of compressors that will be connected to
the ES 6 (maximum 6)
1
# of comp. in
system
‘Remote to Local’
Time before the next compressor is set to Local
control (in case of a Go to local manual action)
20 seconds
20 seconds
‘Unload time’
Time before the next compressor starts running
unloaded (valid in Automatic Operation)
5 seconds
15 seconds
‘Start/Load time’
Time before the next compressor is started (valid
in Automatic Operation)
20 seconds
20 seconds
‘Start/Load Reac. ID1’
Start + Load Reaction time of connection channel
1 Time which the compressor has to run in load
condition after a start command
20 seconds
30 seconds
‘Start/Load Reac. ID2’ Start + Load Reaction time of connection channel 2
20 seconds
30 seconds
‘Start/Load Reac. ID3’ Start + Load Reaction time of connection channel 3
20 seconds
30 seconds
‘Start/Load Reac. ID4’ Start + Load Reaction time of connection channel 4
20 seconds
30 seconds
‘Start/Load Reac. ID5’ Start + Load Reaction time of connection channel 5
20 seconds
30 secondsi
‘Start/Load Reac. ID6’ Start + Load Reaction time of connection channel 6
20 seconds
30 seconds
‘Stop Reaction’
Time which the compressor has to stop after a
stop command
35 seconds
35 seconds
‘Unload Reaction’
Time which the compressor has to run in unload
condition after an unload command
5 seconds
5 seconds
‘Load Reaction’
Time which the compressor has to run in load con-
dition after start command
10 seconds
10 seconds
‘Forced time’
When for a number of days the same compressors
are running in loaded condition, another compres-
sor will be started to force again another compres-
sor to change condition.
7 days
1 day
‘Delta Time’
Maximum allowed difference in running hours
between a stopped and an unloaded compres-
sor. This is to prevent the compressor running
unloaded is always used in favor of the stopped
compressor, although the difference in running
hours is high.
168 hours
168 hours
Appendix A - Central Control System
A.6.1.1 Parameter Listing (Continued)