Cleaning and Storage
119
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
NOTE:
When cleaning the back of the Pump, lightly wipe the battery terminal pins with a
damp cloth. Blot dry immediately (do not wipe dry), using a clean dry lint-free towel.
Do not allow moisture to permeate the terminal pins or the seal.
8. Apply cleaning solution to a lint-free, foam tipped swab (do not use a cotton swab). Blot the
swab onto a dry lint-free towel to remove excess cleaning solution.
9. Wipe down the speaker vent, battery pocket.
10. Ensure that the drainage channel in the battery pocket is clean and clear of any debris.
11. If the keyhole needs cleaning, send the Pump to the Biomed for proper cleaning.
12. Open the Pump’s door using a compatible IV set’s slide clamp.
13. Using a fresh lint-free, foam tipped swab, (do not use a cotton swab), apply cleaning solution
to swab. Blot the swab onto a dry lint-free towel to remove excess cleaning solution.
NOTE:
Do not attempt to clean inside latch holes adjacent to the Direction of Flow label.
NOTE:
When cleaning the tubing channel, do not allow cleaning fluids to seep into or
between the Pump components.
14. Wipe down the surface of door and tubing channel areas.
15. Visually inspect the tubing channel and remove any foreign material. An obstruction in the
tubing channel could cause free flow.
16. Clean the battery module following the steps in “Cleaning the Battery Module” on page 120.
17. Wipe the Pump dry and allow to air-dry in the open air before Pump reuse.
NOTE:
The Pump should remain powered off and the AC Power Adaptor unplugged from a
power source until all cleaning liquids are completely evaporated from the entire
Pump. Allow time for fluids that may have seeped into or between Pump
components to dry. The battery module should remain off the Pump until air drying is
complete.
NOTE:
Allow additional drying time when in cold or humid environments.
18. Dispose of all cleaning materials (including the slide clamp) as required per facility
protocol/biohazard policy.
WARNING
Proper Disposal Required.
To dispose of this device or the associated administration sets,
adhere to local, state, federal and/or other governing regulations.